11. Attachments

The patient-contact materials have undergone biocompatibility testing and have been verified to comply with ISO 10993-1 standards.

warn Use the attachments specified in this chapter. Using other attachments may damage the equipment or result in non-compliance with required specifications. Disposable accessories are not designed for reuse. Reuse may cause complications and affect measurement accuracy.
cautious If storage or use exceeds the specified temperature and humidity range, the accessories may not meet performance specifications. If accessory performance deteriorates due to aging or environmental conditions, please contact only authorized service personnel.

11.1 Treatment of adnexa

name Product Name Model Remark
electrode Preservation difference plus C OBS-DEFP 303A1206 The application section excluding the quota. For adults.
SavePads+CS OBS-DEFP 303A1207 Disposable electrodes for children

Electrode lifespan

All electrode models have a lifespan of up to 48 months, plus a 12-month shelf life. Please replace the electrode before it becomes ineffective. The electrodes are for single use only and cannot be reused. Please replace the electrode immediately after treatment.

Electrode replacement

Note: Do not open the electrode bag during the replacement process!

Press the connector to remove the used or expired electrode and insert the new electrode into the device. See the electrode bag for installation details.

11.2 Battery

name Model Remark
Battery 3c NRL03C 12V, 2.8 Ah, non-rechargeable lithium battery
name Model Remark
3g battery NR103 charger 14.4V, 2.95 Ah, rechargeable lithium-ion battery

Battery replacement

The 3C battery is a non-rechargeable battery with a standby life of 48 months and a shelf life of 12 months. Non-rechargeable batteries should be replaced before the expiration date.

  • The 3G battery is a rechargeable battery with a lifespan of 12 years. It is charged at low capacity and should be replaced before it expires.

For details on battery replacement, please refer to Chapter 6.

11.3 Charging adapter (for rechargeable batteries only)

Compatible with rechargeable batteries (3G). When connected to AC power, the device should be placed in a stable location. To disconnect the device from the power source, disconnect the adapter from the AC power outlet, and then disconnect the adapter from the device. Be sure to close the silicone cap.

name parameter
Rechargeable battery adapter Input 100-240V, 50/60Hz. 0.5A
Output 5.0V∞ 2.0A
warn Battery 3G charging is performed using only the primary medical adapter.